VPAP Adapt SV™ is an adaptive servo-ventilator designed specifically to treat central sleep apnea (CSA) in all its forms, including complex and mixed sleep apnea.
The right support at the right time—adapting breath-by-breath
Unlike conventional SDB therapies such as CPAP, adaptive-servo ventilation:
Peer-reviewed literature shows that adaptive-servo ventilation:
Support when it's needed
The adaptive-servo ventilation algorithm:
Maximizing patient comfort and compliance
VPAP Adapt SV:
Enhanced VPAP Adapt SV
Range of Parameters
The following table lists the ranges of the parameters that can be adjusted in the Settings menu.
Parameter |
Minimum |
Maximum |
Default |
| ASV mode | |||
| EEP (cm H2O)* | 4 |
15 |
5 |
| MIN PS (cm H2O)* | 3 |
6 |
3 |
| MAX PS (cm H2O)* | 9 |
16 |
15 |
| CPAP mode | |||
| CPAP (cm H2O) | 4 |
20 |
10 |
| Start CPAP (cm H2O) | 4 |
20 |
4 |
| Ramp (minutes) | 0 |
45 |
20 |
| Max ramp (minutes) | 0 |
45 |
30 |
*Advanced menu
The VPAP Adapt SV will automatically limit to ensure that:
- MAX PS minus MIN PS is greater than or equal to 5 at all times
- EEP + MAX PS will not exceed 25 cm H2O
To satisfy these limits:
1. Setting EEP automatically adjusts MIN PS then MAX PS
2. Setting MIN PS automatically adjusts MAX PS